• Celignis Analysis Package
    P123 : Composition of Residue from Enzymatic Hydrolysis

Constituents Determined

For every constituent determined via wet-chemical analysis each sample is analysed in duplicate and we provide the compositional data for each replicate along with the average value and the standard deviation between the duplicates. Our analytical protocols typically provide results with a high level of precision, as detailed here.

Examples of the data reports generated can be viewed on the Celignis Database. Please log on to the guest account using email "test@celignis.com" and password "celignis".

Click here to place an order for determining Composition of Residue from Enzymatic Hydrolysis.

Request a QuoteComposition of Residue from Enzymatic Hydrolysis




Analytical Procedure for Composition of Residue from Enzymatic Hydrolysis


☑ Step 1: Acid Hydrolysis of the Sample

When this Method is the first step in the Analysis Package, the acid hydrolysis takes place on the biomass sample with no prior extraction carried out at Celignis. If previous Steps involving the extraction of the sample have take place then the extracted material is used for the acid hydrolysis.

In the case where three different types of extracted material exist (water-extracted, ethanol-extracted, and water- then ethanol-extracted) then the sample that has undergone both water and ethanol extraction is typically used for acid hydrolysis, unless otherwise requested by the customer.

The following steps are involved in the acid hydrolysis of a sample.

1.  The moisture content of the sample is determined, in duplicate.

2.  Approximately 300 mg (with the exact weight noted) of the sample is added to a pressure tube.

3.  3.00 mL of 72% H2SO4 is added by means of an automatic titrator, the weight of the acid added is noted.

4.  The sample is mixed thoroughly with the acid using a glass rod, care is taken that no sample stays adherent to the sides of the tube, but instead stays in contact with the acid.

5.  The tube is transferred to a water bath that is maintained at 30 degrees celcius.

6.  Steps 2-5 are repeated for the duplicate sample.

7.  Every 10 minutes the glass rod for each pressure tube is stirred so that the acid reaches all parts of the sample and complete hydrolysis occurs. This is a crucial step.

8.  Exactly one hour after it is placed in the water bath the pressure tube is removed and placed on a scales and 84 mL of water added (with the weight of the added water recorded). Any acid/sample on the rod is removed from the rod at this point using this water.

9.  A lid is screwed on the tube and the tube is inverted several times to ensure thorough mixing of the acid.

10. Two sugar recovery solution (SRS) pressure tubes are prepared in order to monitor the sugar-loss associated with the second-stage-hydrolysis. This involves the following steps:

(a) 348 microlitres of 72% H2SO4 is added to a test tube containing a solution containing a known weight (approximately 10 g) of a sugar standard. This standard should be of a similar sugar composition to that expected of the samples being analysed. The acid and sugar solution are thoroughly mixed.
(b) The sugar-acid mixture is transferred to a pressure tube which is then sealed.

11. All SRS and sample pressure tubes are placed in an autoclave which is run at 121 degrees celcius for 60 minutes.

12. After the temperature in the autoclave drops to under 80C the tubes are removed and are left (closed) in the lab until they reach room temperature.

13. The hydrolysates are then filtered (using vacuum suction) through filter crucibles of known weight and the resulting filtrate is stored.

14. Any residual solids are washed out from the tube using deionised water until all the Acid Insoluble Residue resides on the filter crucible.

☑ Step 2: Determination of ASL Using UV-Vis Spectroscopy

The following steps are involved in the determination of the Acid Soluble Lignin (ASL) content:

1. The hydrolysate (filtrate) from the Acid Hydrolysis Stage, is quantitatively transferred using a pipette to a 1 cm path-length (3 mL volume) quartz cuvette.

2. The UV-Visible (190-520nm) transmission spectrum of the sample is then obtained using the HP 8452A diode-array spectrophotometer.

3. If the absorbance at a given wavelength is outside of the linear region (considered to be between 0.7 and 1.0 absorbance units) then the sample is diluted with water until the value falls within this region.

3. This dilution and UV-Visible analysis is carried out twice for each hydrolysate, meaning that each sample will have four spectra collected (two spectra for each of the duplicate hydrolysates). 4. The Acid Soluble Lignin content is then calculated based on the absorbance value at 205nm, the dilution factor, and a given absorptivity constant.

☑ Step 3: Hydrolysate Dilution and Storage

Following the Acid Hydrolysis Step, the hydrolysate (the filtrate from the vacuum filtration) is diluted by a factor of 5 using a fucose-in-water solution. Fucose is the internal standard that is used in the chromatographic analysis of the hydrolysate. Following this dilution the samples are either immediately put on the chromatography system or stored in a freezer for future analysis.

☑ Step 4: Gravimetric Determination of Klason Lignin

The following steps are involved in the gravimetric determination of the Klason Lignin, Acid Insoluble Residue, and Acid Insoluble Ash contents:

1.  The filter crucibles from the Acid Hydrolysis Step are placed in an oven overnight and dried at 105C.

2.  These crucibles are then weighed to determine the Acid Insoluble Residue content.

3.  The crucibles are then placed in a Nabertherm L-240H1SN muffle furnace and the following program used:

(i)     Ramp from room temperature to 105C.
(ii)    Hold at 105C for 12 minutes.
(iii)   Ramp to 250C at 10C/minute.
(iv)   Hold at 250C for 30 minutes.
(v)    Ramp to 575C at 20C/ minute.
(vi)   Hold at 575C for 180 minutes.
(vii)  Allow temperature to drop to 105C.
(viii) Hold at 105C until the samples are removed.

4.  The fiter crucibles are then weighed and the Acid Insoluble Residue, and Acid Insoluble Ash content calculated.

5.  The Klason Lignin content is then calculated by subtracting the Acid Insoluble Ash content from the Acid Insoluble Residue content.

☑ Step 5: Chromatographic Analysis of the Hydrolysate

Equipment Used for Composition of Residue from Enzymatic Hydrolysis Analysis



Ion Chromatography

A Dionex ICS-3000 system that is equipmed with electrochemical, conductivity, and ultraviolet-visible detectors.



UV-Vis Spectrophotometer

A HP 8452A Ultraviolet-Visible spectroscopy device that is used for the determination of the acid soluble lignin concentration of the acid hydrolysate of samples.



Muffle Furnace

A Nabertherm furnace is used for the determination of the ash content of samples and also in the analytical protocol for determining Klason lignin content.



Autoclave

An autoclave is used in the protocols for the determination of the lignin and structural sugars (cellulosic and hemicellulosic) contents of samples.



...